Test Code ZN_S Zinc, Serum
Additional Codes
MAYO ID: ZN_S
Meditech Mnemonic:
ZINC
Specimen Required
Patient Preparation: High concentrations of gadolinium, iodine, and barium are known to interfere with most metal tests. If gadolinium-, iodine, or barium-containing contrast media has been administered, the specimen should not be collected for 96 hours.
Supplies:
-Metal Free Specimen Vial (T173)
-Metal Free B-D Tube (No Additive), 6 mL (T184)
Collection Container/Tube: 6-mL Plain, royal blue-top Vacutainer plastic trace element blood collection tube
Submission Container/Tube: 7-mL Mayo metal-free, screw-capped, polypropylene vial
Specimen Volume: 0.8 mL
Collection Instructions:
1. This specimen must always be drawn first.
2. Do not collect specimen from a line.
3. Allow the specimen to clot for 30 minutes; then centrifuge the specimen to separate serum from the cellular fraction. Serum must be removed from cellular fraction within 4 hours of specimen collection. Avoid hemolysis.
4. Remove the stopper. Carefully pour specimen into a Mayo metal-free, polypropylene vial, avoiding transfer of the cellular components of blood. Do not insert a pipet into the serum to accomplish transfer, and do not ream the specimen with a wooden stick to assist with serum transfer.
5. See Metals Analysis Specimen Collection and Transport for complete instructions.
Useful For
Detecting zinc deficiency
Special Instructions
Method Name
Dynamic Reaction Cell-Inductively Coupled Plasma-Mass Spectrometry (DRC-ICP-MS)
Reporting Name
Zinc, SSpecimen Type
SerumSpecimen Minimum Volume
0.2 mL
Specimen Stability Information
Specimen Type | Temperature | Time | Special Container |
---|---|---|---|
Serum | Refrigerated (preferred) | 28 days | METAL FREE |
Ambient | 28 days | METAL FREE | |
Frozen | 28 days | METAL FREE |
Reject Due To
Gross hemolysis | Reject |
Gross lipemia | OK |
Gross icterus | OK |
Reference Values
0-10 years: 60-120 mcg/dL
11-17 years: 66-110 mcg/dL
≥18 years: 60-106 mcg/dL
Day(s) Performed
Monday through Saturday
Report Available
1 to 3 daysPerforming Laboratory
Mayo Clinic Laboratories in RochesterTest Classification
This test was developed and its performance characteristics determined by Mayo Clinic in a manner consistent with CLIA requirements. It has not been cleared or approved by the US Food and Drug Administration.CPT Code Information
84630
LOINC Code Information
Test ID | Test Order Name | Order LOINC Value |
---|---|---|
ZN_S | Zinc, S | 5763-8 |
Result ID | Test Result Name | Result LOINC Value |
---|---|---|
7735 | Zinc, S | 5763-8 |
Forms
If not ordering electronically, complete, print, and send General Test Request (T239) with the specimen.